The electronic balance occasionally has some small faults when in use, so how to deal with and solve the faults:
After each weighing of the electronic balance, the indication value does not return to zero
The cause of the failure and the treatment method:
1. The balance has a short warm-up time, so it should be warmed up for more than 30 minutes.
2. The balance is not level, adjust the level of the balance.
3. The balance should be calibrated regularly.
4. The linearity error is too large and exceeds the allowable range. Linear adjustment should be made according to the balance manual.
No display at all after the electronic balance is turned on
The cause of the failure and the treatment method:
1. The parameters in the balance menu are not set properly: After entering the menu, use the "RESET" function to exit the menu correctly and return to the factory settings
2. The environment where the balance is placed is too poor: Environmental factors include: vibration, airflow, temperature, and external magnetic field. The above environment must be improved and the windshield window of the weighing room must be closed for the balance to work normally.
Electronic balance displays "L"
The cause of the failure and the treatment method:
Check whether the weighing pan is placed properly, the one-thousandth balance is unreliable and the spring is too tight, so it can be loosened. Check whether the connectors are loose, check whether the position of the zero detection grating changes, and adjust the grating position up and down.