The related substances test of Paracetamol API is given in European Pharmacopoeia monograph. In this method, it was mentioned to use ‘end-capped solid core octadecylsilyl silica gel, 0.10m x 2.1 mm; 2.7 µm’ HPLC column. This method equivalency study has been carried out for related substances of Paracetamol API as per EP monograph to establish equivalency between Welch Boltimate® Core-shell HPLC column and Waters Cortecs® HPLC Column . This report presents and discusses the results obtained in the equivalency study.
Column
Welch Boltimate ® Core-shell, LP-C18 (100 x 2.1)mm, 2.7µm
Waters CORTECS® C18 (100 x 2.1)mm, 2.7µm
Chromatographic system
PreColumn | End-capped solid core octadecylsilyl silca gel (0.005m x 2.1mm, 2.7µm) |
Column | 1. Welch Boltimate® LP C18 (100 x 2.1)mm, 2.7µm 2. Waters Cortecs® C18 (100 x2.1)mm, 2.7µm |
Wavelength | 254nm |
Flow rate | 0.3mL/minute |
Injection Volume | 5 µL |
Column temperature | 30°C |
Auto sampler temperature | 25°C |
Results
1. Welch Boltimate® LP C18 (100 x 2.1)mm, 2.7µm
2. Waters Cortecs® C18 (100 x2.1)mm, 2.7µm
Parameter | Column-1 | Column-2 |
The Resolution between the peak due to impurity K and paracetamol | 10.3 | 23.0 |
Name | Relative Retention Time (As per EP) |
Relative Retention Time (Column-1) |
Relative Retention Time (Column-2) |
Impurity K | 0.4 | 0.40 | 0.38 |
Impurity J | 10.1 | 8.17 | 9.34 |
Conclusion
From the results of equivalency study, the equivalency between the Welch Boltimate® HPLC column and Waters Cortecs® HPLC column is proved. Hence a Welch Boltimate® HPLC Column can be used for related substances by HPLC of Paracetamol API as per EP monograph.
Chromatograph of Solvent mixture
Fig 1: Welch Boltimate® LP C18 (100 x 2.1)mm, 2.7µm
Fig 2: Waters Cortecs® C18 (100 x2.1)mm, 2.7µm
Chromatograph of Reference solution (a)
Fig 3: Welch Boltimate® LP C18 (100 x 2.1)mm, 2.7µm
Fig 4: Waters Cortecs® C18 (100 x2.1)mm, 2.7µm
Chromatograph of Reference solution (b)
Fig 5: Welch Boltimate® LP C18 (100 x 2.1)mm, 2.7µm
Fig 6: Waters Cortecs® C18 (100 x2.1)mm, 2.7µm
Chromatograph of Reference solution (c)
Chromatograph of Test Solution
Fig 9 :Welch Boltimate® LP C18 (100 x 2.1)mm, 2.7µm
Fig 10: Waters Cortecs® C18 (100 x2.1)mm, 2.7µm
View PDF: https://www.welch-us.com/wp-content/uploads/2021/08/Paracetamol_Method-Equivalency-report-11.pdf
If you have any problem or require further information, please contact [email protected].
Comparison Between Welch Xtimate® HPLC & A Well-known HPLC Column By Azithromycin API As Per USP.
May 22,2023Thanks Everyone For Visiting Our Booth At Eastern Analytical Symposium Exhibition
May 22,2023Your email address will not be published. Required fields are marked *