Chiral chromatographic columns are generally used for the separation of chiral isomers. In addition to chiral isomers, they also have certain effects on the separation of other isomers or components with very similar structures.
Applications
Separation of p-cyano benzyl chloride & m-cyano benzyl chloride
The structure is shown in the figure:
p-cyanobenzyl chloride m-cyano benzyl chloride
The two substances are meta- and para-isomers, and there is no separation trend in the reversed-phase system with different selectivity chromatographic columns, so they cannot be separated. Finally, a chiral chromatographic column was used to achieve separation.
Chromatographic conditions
Chromatographic column: Ultisil® Cellu-D 4.6*250mm, 5μm.
Mobile phase: n-hexane/isopropanol=98/2.